Computer-aided Analysis of Radial Head Morphometry

Paul Puchwein, MD; Nima Heidari, MD; Katrin Dorr, MD; Lukas Struger; Wolfgang Pichler, MD

Abstract

Operative treatment of displaced and comminuted radial head fractures involves internal fixation with plates and screws in cases where reconstruction is possible and replacement with a radial head prosthesis when comminution renders the radial head unreconstructable. The purposes of this study were to evaluate the morphometry of the radial head using a modern technique and to compare the findings with several commercially available radial head prostheses.

Computed tomography scans of 30 cadaveric elbows and 3-dimensional reconstructions were used to analyze the morphometry of the proximal radius. Results were compared with the manufacturer data of several radial head prostheses. Mean diameter of the radial head at the level of the fovea was 19±1.58 mm (range, 15.82–21.81 mm) in the anteroposterior plane and 18.62±1.78 mm (range, 15.48–22.21 mm) in the radioulnar plane. Mean diameter of the radial head at its widest part was 23.15±1.94 mm (range, 19.45–26.49 mm) in the anteroposterior plane and 22.44±1.73 mm (range, 19.64–25.44 mm) in the radioulnar plane. Mean diameter of the radial head at the level of the head–neck junction was 15.42±1.59 mm (range, 11.80–18.46 mm) in the anteroposterior plane and 14.75±1.39 mm (range, 12.32–17.31 mm) in the radioulnar plane. Statistically significant sex differences existed in the maximum diameter of the radial head, the diameter at the level of the head–neck junction, and the length of the radial head. Currently available radial head prostheses cover the range of sizes encountered. Products with a choice of head and stem sizes in any combination are preferable. In unstable elbow fractures, correct implant size is an important factor to avoid subluxation of the radial head (Mason type IV fractures) if collateral ligaments are sufficient.

The authors are from the Department of Traumatology (PP, LS, WP) and the Department of Radiology (KD), Medical University of Graz, Graz, Austria; and from the Specialist Registrar Royal London Rotation (NH), Royal London Hospital, London, United Kingdom.

The authors have no relevant financial relationships to disclose.

Correspondence should be addressed to: Paul Puchwein, MD, Department of Traumatology, Medical University of Graz, Auenbruggerplatz 7a, A-8036 Graz, Austria (paul.puchwein@medunigraz.at).

Abstract

Operative treatment of displaced and comminuted radial head fractures involves internal fixation with plates and screws in cases where reconstruction is possible and replacement with a radial head prosthesis when comminution renders the radial head unreconstructable. The purposes of this study were to evaluate the morphometry of the radial head using a modern technique and to compare the findings with several commercially available radial head prostheses.

Computed tomography scans of 30 cadaveric elbows and 3-dimensional reconstructions were used to analyze the morphometry of the proximal radius. Results were compared with the manufacturer data of several radial head prostheses. Mean diameter of the radial head at the level of the fovea was 19±1.58 mm (range, 15.82–21.81 mm) in the anteroposterior plane and 18.62±1.78 mm (range, 15.48–22.21 mm) in the radioulnar plane. Mean diameter of the radial head at its widest part was 23.15±1.94 mm (range, 19.45–26.49 mm) in the anteroposterior plane and 22.44±1.73 mm (range, 19.64–25.44 mm) in the radioulnar plane. Mean diameter of the radial head at the level of the head–neck junction was 15.42±1.59 mm (range, 11.80–18.46 mm) in the anteroposterior plane and 14.75±1.39 mm (range, 12.32–17.31 mm) in the radioulnar plane. Statistically significant sex differences existed in the maximum diameter of the radial head, the diameter at the level of the head–neck junction, and the length of the radial head. Currently available radial head prostheses cover the range of sizes encountered. Products with a choice of head and stem sizes in any combination are preferable. In unstable elbow fractures, correct implant size is an important factor to avoid subluxation of the radial head (Mason type IV fractures) if collateral ligaments are sufficient.

The authors are from the Department of Traumatology (PP, LS, WP) and the Department of Radiology (KD), Medical University of Graz, Graz, Austria; and from the Specialist Registrar Royal London Rotation (NH), Royal London Hospital, London, United Kingdom.

The authors have no relevant financial relationships to disclose.

Correspondence should be addressed to: Paul Puchwein, MD, Department of Traumatology, Medical University of Graz, Auenbruggerplatz 7a, A-8036 Graz, Austria (paul.puchwein@medunigraz.at).

Fractures of the radial head and neck are one of the most common fractures about the elbow, comprising 33% of all elbow fractures and 1.5% to 4% of all fractures in adults.1 A common mechanism of injury is a fall on an outstretched hand with an extended and pronated forearm.2 Operative treatment of displaced and comminuted radial head fractures involves open reduction and internal fixation with plates and screws in cases where reconstruction is possible and replacement with a radial head prosthesis when comminution renders the radial head unreconstructable.3

In 1954, Mason4 classified the displacement of radial head fractures. This classification was subsequently modified by Broberg and Morrey5 in 1987 and Hotchkiss6 in 1997. In Hotchkiss type III fractures, resection or replacement of the radial head with a prosthesis may be necessary. However, these implants may not always precisely recreate the complex anatomy of the proximal radius.7,8

The purpose of this study was to evaluate the morphometry of the radial head using a modern technique and compare the findings with several commercially available radial head prostheses.

Materials and Methods

Thirty cadaveric elbows preserved using Thiel’s embalming method were examined in the supine position using a 64-slice Siemens Somatom Sensation computed tomography system (0.6 mm contiguous axial slices) (Siemens Medical Solutions USA, Inc, Malvern, Pennsylvania). Digital Imaging and Communications in Medicine raw data sets were reconstructed using Mimics 3-dimensional software (Materialise, Leuven, Belgium). Extremities with severe arthrosis, evidence of trauma, or other pathological changes were excluded from the study. Mean age of the donors at death was 72.5 years (range, 59–98 years). Fourteen elbows were from male donors and 16 were from female donors. Sixteen elbows were right elbows and 14 were left elbows. The results were recorded on Excel 2003 software (Microsoft Corp, Redmond, Washington). Two-sample t test was used for calculating sex differences. A P value of .05 or less was considered statistically significant.

To the authors’ knowledge, no published standardized methods for measuring the radial head exist. Therefore, the following points of reference were used to describe the geometry of the radial head and neck for the purposes of this study.

The first 4 points were set on the most proximal projections of the fovea of the radial head to measure the anteroposterior (AP) and radioulnar (RU) diameters (Figure 1). The radial head is slightly barrel shaped and widens before narrowing again to join the radial neck. Measurements were performed in the AP and RU planes between points set at the widest part of the radial head to quantify its maximum diameters (Figure 2). Four more points were set at the head–neck junction as determined by the vertex of the curve of the neck as it joins the shaft in the AP and RU planes. The vertex was defined by the point of the curve that is farthest from a line between the beginning and end of the curve (Figure 3).

Screenshots of Mimics 3-dimensional software (Materialise, Leuven, Belgium) showing measurements of the most proximal projections of the fovea. Coronal (A), transverse (B), and sagittal (C) views of the radial head of the left elbow (A). Anteroposterior 3-dimensional reconstruction of the elbow (D).

Figure 1: Screenshots of Mimics 3-dimensional software (Materialise, Leuven, Belgium) showing measurements of the most proximal projections of the fovea. Coronal (A), transverse (B), and sagittal (C) views of the radial head of the left elbow (A). Anteroposterior 3-dimensional reconstruction of the elbow (D).

Screenshots of Mimics 3-dimensional software (Materialise, Leuven, Belgium) showing measurements of the widest part of the radial head and junction area. Coronal view of the radial head of the left elbow (A). Transverse view of the proximal radial head (B). Coronal view of the radial head of the left elbow (C). Transverse view of the radial head at the level of the junction (D).

Figure 2: Screenshots of Mimics 3-dimensional software (Materialise, Leuven, Belgium) showing measurements of the widest part of the radial head and junction area. Coronal view of the radial head of the left elbow (A). Transverse view of the proximal radial head (B). Coronal view of the radial head of the left elbow (C). Transverse view of the radial head at the level of the junction (D).

Screenshots of Mimics 3-dimensional software (Materialise, Leuven, Belgium) showing 1 of the 4 shaft–head angles (radial angle) and measurement of the head length (ulnar head length). Coronal view of the radial head of the left elbow showing the radial and ulnar shaft points measured 5 mm distal to corresponding junction points (A). Sagittal view showing the anterior and posterior junction and shaft points (B). Oblique front view on the reconstructed elbow showing the radial shaft–head angle (C). Coronal view of the radial head showing the ulnar head length between the fovea point and the junction point (D).

Figure 3: Screenshots of Mimics 3-dimensional software (Materialise, Leuven, Belgium) showing 1 of the 4 shaft–head angles (radial angle) and measurement of the head length (ulnar head length). Coronal view of the radial head of the left elbow showing the radial and ulnar shaft points measured 5 mm distal to corresponding junction points (A). Sagittal view showing the anterior and posterior junction and shaft points (B). Oblique front view on the reconstructed elbow showing the radial shaft–head angle (C). Coronal view of the radial head showing the ulnar head length between the fovea point and the junction point (D).

Two pairs of points 5 mm distal to those at the head–neck junction on the periosteal surface of the radial shaft in the AP and RU planes were chosen and labeled shaft (Figure 3). The head–shaft angle was calculated between the points max head, junction, and shaft at each of the measurement points for the anterior, posterior, ulnar, and radial sides (Figure 3).

To determine the lengths of the anterior, posterior, radial, and ulnar aspects of the radial head, the distance between each of the 4 fovea points and the corresponding junction point was measured (Figure 3). To quantify the flexion between the radial head and neck in the sagittal plane, a circle was constructed using 3 points on the coverture. The radius of this constructed circle was measured (Figure 4). The dimensions of various radial head prostheses are shown in Table 1.

Screenshot of Mimics 3-dimensional software (Materialise, Leuven, Belgium) showing a circle quantifying the flexion between the radial head and neck in the sagittal plane.

Figure 4: Screenshot of Mimics 3-dimensional software (Materialise, Leuven, Belgium) showing a circle quantifying the flexion between the radial head and neck in the sagittal plane.

Dimensions and Specifications of Radial Head Prostheses

Table 1: Dimensions and Specifications of Radial Head Prostheses

Results

Mean diameter of the radial head at the level of the fovea was 19±1.58 mm (range, 15.82–21.81 mm) in the AP plane and 18.62±1.78 mm (range, 15.48–22.21 mm) in RU plane. Mean diameter of the radial head at its widest part was 23.15±1.94 mm (range, 19.45–26.49 mm) in the AP plane and 22.44±1.73 mm (range, 19.64–25.44 mm) in the RU plane. Mean diameter at the level of the head–neck junction was 15.42±1.59 mm (range, 11.80–18.46 mm) in the AP plane and 14.75±1.39 mm (range, 12.32–17.31 mm) in the RU plane.

Because the differences between the mean AP and RU diameters were marginal, the radial head was assumed to be circular, mean AP and RU measurements at each level (labeled diameter) were used. Mean diameter of the fovea was 19.15±1.54 mm in male donors and 18.50±1.66 mm in female donors. This difference did not reach statistical significance (P=.28). However, for the widest part of the radial head, sex-dependent differences were observed (men, 23.81±1.58 mm; women, 21.90±1.37 mm; P<.01). Similar results were observed at the head–neck junction (men, 15.76±1.39 mm; women, 14.49±1.12 mm; P=.01). No statistically significant difference existed between the right and left radial head measurements (P=.49).

Mean radial head length on each side (anterior, 11.81±1.57 mm; posterior, 11.77±1.41 mm; radial, 11.89±1.52 mm; ulnar, 11.72±1.42 mm) was compared between male and female donors, resulting in significant differences (men, 12.41±1.7 mm; women, 11.26±0.83 mm; P<.01). Mean radius of the curvature was 8.76±2.50 mm, with no significant difference between the sexes (P=.59). Mean angle between the radial head and shaft was 158.8° (Table 2).

Measurement Results

Table 2: Measurement Results

Comparing this study’s results with the dimensions of commercially available radial head prostheses is demanding because different controversial concepts exist regarding radial head displacement.

The Acumed Anatomical Radial Head System (Acumed, Hillsboro, Oregon) is a metal implant with a free combinable head and cementless stem components. Head diameter options are 20, 22, 24, 26, and 28 mm, which is more than the average head diameter in the current study. With a preset 4° ventral and 4° ulnar angulation, right and left implants are needed.

Ascension Orthopedics (Austin, Texas) offers a metal modular radial head implant and a pyrocarbon radial head implant with 3 head sizes (20, 22, and 24 mm) and 4 stem sizes (each with a standard or a long collar) that can be implanted without cement.

The Liverpool Radial Head Replacement (Biomet, Inc, Warsaw, Indiana) has a shaft angulation of 10°, two head diameters, and 13 different stems. The stems are bond-coated, which should provide a better fixation. The ExploR Radial Head by Biomet can be cemented or press-fit. Three different head sizes (20, 22, and 24 mm), 5 different head lengths, and 5 different stems are offered.

The Radial Head metal (CoCrMo) implant (Corin Group PLC, Cirencester, United Kingdom) is coated with hydroxyapatite and titanium. Three head sizes (18, 21, and 25 mm) and 3 different stems are offered.

Another radial head prosthesis is the rHead implant (Small Bone Innovations, Inc, Morrisville, Pennsylvania), which is offered in a lateral version (head can be loaded from lateral using a less invasive approach) and a recon version (bipolar head with the ability of a free angulation up to 20°). The implants are made of metal (CoCr) and feature a fully coated stem. The shaft angulation is 12°. Three head sizes (18, 21, and 24 mm) and 4 stems sizes with 2 collar lengths are available.

The Solar Radial Head implant by Stryker (Kalamazoo, Michigan) has a cemented stem and features 2 head sizes with 5 different stems.

Tornier (Saint-Ismier Cedex, France) offers different radial head prostheses. The commonly used CRF II implant was introduced approximately 2 decades ago. It features a shaft angulation of 15°, a head range of motion of 35°, and is made of metal (CrCo). Two head sizes (19 and 22 mm) and 2 cementable stems (6.5 and 8 mm) are available. The newer MoPyC implant is made of pyrocarbon and offers the same shaft angulation and a cement-less press-fit implantation. Its smaller head sizes (18, 20, and 22 mm) are comparable with the sizes found in the current study. Four neck sizes and 4 shaft sizes are free combinable.

Wright Medical Technology, Inc (Arlington, Tennessee), offers 2 radial head implants. The Swanson Titanium Radial Head Implant is available in 5 head diameters (19–23 mm). The stem size depends on the head, and the components are not combinable. A Silicon Head version is also offered. The modular Evolve system is offered with 6 head diameters (18, 20, 22, 24, 26, and 28 mm), lengths of 11, 13, or 15 mm, and 6 stem sizes, each with a standard neck, a 2-mm neck, and a 4-mm neck. Many combinations are possible.

Discussion

Operative treatment of radial head and neck fractures consists of open reduction and internal fixation when a stable, reliable fixation can be achieved. In fractures where comminution renders the fracture unreconstructable, the fragments are excised and a metal radial head prosthesis is inserted.

Some biomechanical studies have emphasized the importance of correctly sizing the radial head prosthesis at the time of implantation.1 A short, small prosthesis can lead to valgus instability, whereas a long implant has the potential to overstuff the radiocapitellar joint. Some authors have suggested that a difference of no more than 2 mm from the patient’s native anatomy is tolerated.9

In unstable elbow fractures, correct implant size is an important factor to avoid subluxation of the radial head (Mason type IV fractures) if collateral ligaments are sufficient. In the current authors’ opinion, posterior subluxation can occur in cases of insufficient medial collateral ligaments.

Tejwani and Mehta1 reported that 60% of the axial load from the forearm is transmitted through the radiocapitellar joint, particularly with the elbow flexed and the forearm pronated. The highest loads were recorded with the elbow extended and forearm pronated.1 Itamura et al10 reported discrepancies between the anatomy of the radial head and available radial head prostheses. Beredjiklian et al7 used magnetic resonance imaging to investigate the morphology of the radial head and reported results similar to those of the current study. They measured the minimum (22 mm) and maximum (23 mm) diameter of the radial head at the level of the convexity, comparable with the maximum diameter of the radial head in the AP (23.15 mm) and RU (22.44 mm) planes.7 Mean radial head length (12 mm) in the study by Beredjiklian et al7 was similar to the mean dorsal, ventral, radial, and ulnar length (11.8±1.19 mm) in the current study.

Using the Poli SKY II coordinate measuring machine (Quality Control Technology Ltd, Nottingham, United Kingdom) and ProENGINEER software (PTC, Needham, Massachusetts), Swieszkowski et al11 examined 16 radial heads and reported a mean diameter of 23.36±1.14 mm. In a biometric study, Captier et al12 measured 96 radial bones using a vernier caliper and goniometer and found the fovea to be elliptical (more than 1 mm difference between AP and RU diameters) in 57% of bones; mean AP diameter was 21.6±2.9 mm and mean RU diameter was 21±2.7 mm. Compared with the current study’s results, Captier et al12 measured a slightly smaller diameter in both planes. They also reported a mean radial head length of 15.3 mm,12 whereas in the current study it was 11.80 mm. This may be due to the use of dry bones as opposed to fresh. Captier et al12 detected a relationship between the angle of the axis of the radial neck and diaphysis and the shape of the fovea. They reported a mean angle of 166.75°±3° in specimens with a circular fovea and 168.62°±3.2° in specimens with an oval-shaped fovea. They postulated that pronation is facilitated due to deviation of the radial tuberosity by 2 mm from the incisura.12

Popovic et al8 scanned 51 proximal radii using computed tomography and compared their measurements with those of a commercially available radial head prosthesis. They reported results for radial head diameter (22.8±1.9 mm maximum and 21.8±1.9 mm minimum) comparable to the current study and described sex-dependent differences in radial head size. They showed that the smaller 19-mm implant is too small for most adults, whereas the 22-mm implant should fit in most cases. They also recommended that the shaft and head of the implant should be used in any combination because no correlation seemed to exist between head size and intramedullary neck diameter.8

Itamura et al10 examined 22 elbows using computed tomography with the forearm placed in different positions and reported a maximum diameter of 22.3±2.4 mm and a minimum diameter of 20.9±1.9 mm, comparable with the measurements in the current study (Table 3). However, their mean radial head length measurement (11.8±1.19 mm) differed from those of the current study because different measurement points were used.

Measurement Results in Comparable Studies

Table 3: Measurement Results in Comparable Studies

Numerous radial head prostheses from a variety of manufacturers are available on the market (Table 1). Most implants come in several head sizes covering the anatomical range demonstrated by the current study’s measurements, but not all products offer the possibility of fully interchangeable head and shaft sizes.

New isoelastic (pyrocarbon) prostheses should offer better gliding ability without wear. First results from animal studies and retrospective clinical studies show advantages of pyrocarbon prostheses compared with metal prostheses.13–16 Recent case series on the use of bipolar radial head prostheses report good functional results.17,18 In a cadaveric study, Moon et al19 reported that a bipolar prosthesis was unable to resist subluxation as well as the native radial head or a monopolar prosthesis.

Summarizing the aforementioned studies, reliable, comparable values were found for small and large radial head diameters, whereas radial head and neck length and head–diaphysis angle measurements are difficult to compare in the absence of well-defined landmarks.

In unstable elbow fractures, correct implant size is an important factor to avoid subluxation of the radial head (Mason type IV fractures) if collateral ligaments are sufficient. Measurement of the contralateral side is a valid option to determine the size of the fractured radial head and to choose the correct implant size. Currently available radial head prostheses cover the range of sizes encountered in the current study. Implants with a choice of head and stem sizes in any combination are preferable. Several prosthesis design features cannot be easily compared, but anatomic ovoid implants affect kinematics more physiologically.

References

  1. Tejwani NC, Mehta H. Fractures of the radial head and neck: current concepts in management. J Am Acad Orthop Surg. 2007; 15(7):380–387.
  2. Gebauer M, Rücker AH, Barvencik F, Rueger JM. Therapie der Radiusköpfchenfraktur. Unfallchirurg. 2005; 108(8):657–667.
  3. Pike JM, Athwal GS, Faber KJ, King GJ. Radial head fractures—an update. J Hand Surg Am. 2009; 34(3):557–565. doi:10.1016/j.jhsa.2008.12.024 [CrossRef]
  4. Mason ML. Some observations on fractures of the head of the radius with a review of one hundred cases. Br J Surg. 1954; 42(172):123–132. doi:10.1002/bjs.18004217203 [CrossRef]
  5. Broberg MA, Morrey BF. Results of treatment of fracture-dislocations of the elbow. Clin Orthop Relat Res. 1987; (216):109–119.
  6. Hotchkiss RN. Displaced fractures of the radial head: internal fixation or excision?J Am Acad Orthop Surg. 1997; (5):1–10.
  7. Beredjiklian PK, Nalbantoglu U, Potter HG, Hotchkiss RN. Prosthetic radial head components and proximal radial morphology: A mismatch. J Shoulder Elbow Surg. 1999; 8(5):471–475. doi:10.1016/S1058-2746(99)90079-4 [CrossRef]
  8. Popovic N, Djekic J, Lemaire R, Gillet P. A comparative study between proximal radial morphology and the floating radial head prosthesis. J Shoulder Elbow Surg. 2005; 14(4):433–440. doi:10.1016/j.jse.2004.10.012 [CrossRef]
  9. Moghaddam A, Lennert A, Studier-Fischer S, Wentzensen A, Zimmermann G. Prosthesis after comminuted radial head fractures Midterm results. Unfallchirurg. 2008; 111(12):997–1004. doi:10.1007/s00113-008-1514-9 [CrossRef]
  10. Itamura JM, Roidis NT, Chong AK, Vaishnav S, Papadakis SA, Zalavras C. Computed tomography study of radial head morphology. J Shoulder Elbow Surg. 2008; 17(2):347–354. doi:10.1016/j.jse.2007.07.019 [CrossRef]
  11. Swieszkowski W, Skalski K, Pomianowski S, Kedzior K. The anatomic features of the radial head and their implication for prosthesis design. Clin Biomech (Bristol, Avon). 2001; 16(10):880–887. doi:10.1016/S0268-0033(01)00075-4 [CrossRef]
  12. Captier G, Canovas F, Mercier N, Thomas E, Bonnel F. Biometry of the radial head: biomechanical implications in pronation and supination. Surg Radiol Anat. 2002; 24(5):295–301. doi:10.1007/s00276-002-0059-9 [CrossRef]
  13. Cook SD, Beckenbaugh RD, Redondo J, Popich LS, Klawitter JJ, Linscheid RL. Long-term follow-up of pyrolytic carbon metacarpophalangeal implants. J Bone Joint Surg Am. 1999; 81(5):635–648.
  14. Cook SD, Thomas KA, Kester MA. Wear characteristics of the canine acetabulum against different femoral prostheses. J Bone Joint Surg Br. 1989; 71(2):189–197.
  15. Sarris IK, Kyrkos MJ, Galanis NN, Papavasiliou KA, Sayegh FE, Kapetanos GA. Radial head replacement with the MoPyC pyrocarbon prosthesis. J Shoulder Elbow Surg. 2012; 21(9):1222–1228. doi:10.1016/j.jse.2011.12.004 [CrossRef]
  16. Lamas C, Castellanos J, Proubasta I, Dominguez E. Comminuted radial head fractures treated with pyrocarbon prosthetic replacement. Hand (N Y). 2011; 6(1):27–33. doi:10.1007/s11552-010-9282-8 [CrossRef]
  17. Burkhart KJ, Mattyasovszky SG, Runkel M, et al. Mid- to long-term results after bipolar radial head arthroplasty. J Shoulder Elbow Surg. 2010; 19(7):965–972. doi:10.1016/j.jse.2010.05.022 [CrossRef]
  18. Celli A, Modena F, Celli L. The acute bipolar radial head replacement for isolated unreconstructable fractures of the radial head. Musculoskelet Surg. 2010; 94(suppl 1):S3–S9. doi:10.1007/s12306-010-0062-y [CrossRef]
  19. Moon JG, Berglund LJ, Zachary D, An KN, O’Driscoll SW. Radiocapitellar joint stability with bipolar versus monopolar radial head prostheses. J Shoulder Elbow Surg. 2009; 18(5):779–784. doi:10.1016/j.jse.2009.02.011 [CrossRef]

Dimensions and Specifications of Radial Head Prostheses

ProsthesisDiameter (Length), mm
Special Features
HeadStem
Anatomic Radial Head Systema206 (25)Chirality (different implants for right/left arms); head-socket +0, +2, +4, +6, and +8
227 (25)
248 (25)
269 (25)
2810 (25)
Carbon Modular Radial Headb204 stem sizes each in long versionPyrocarbon
22
24
ExploR Radial Headc205 stem sizes
22
24 (10, 12, 14, 16, 18)
Liverpool Radial Head Replacementc16(6–18, step 2)10° angulation; bond-coated stem
18(14–24, step 2)
Radial Head metal (CoCrMo)d18 (9, 13)10 (28)
21 (11,15)12 (31)
25 (12,17)14 (33)
rHeade18 (7.2)6.4 (16)Recon version
21 (10.2)7.2 (18)
24 (13.2)8.0 (20)
8.8 (22)
Solar Radial HeadfSmall8
Medium9
11
12
15
CRF II Judetg196.5Bipolar
228
MoPyCg18 (9)7 (40)Pyrocarbon neck; 15° angulation; neck lengths: 5, 6.5, 8, 10 mm
20 (10)8 (44)
22 (11)9 (48)
10 (50)
Evolve Prolineh184.5–9.5+2 and +4 mm
20Step 1head
22+2 and +4
24
26
28
Swansonh1927–32Not modular combinable
20
21
22
23 (10–13.5)

Measurement Results

MeasurementMean±SDRange
Diameter
  AP19.00±1.58 mm15.82–21.81 mm
  RU18.62±1.78 mm15.48–22.21 mm
Maximum head
  AP23.15±1.94 mm19.45–26.49 mm
  RU22.44±1.73 mm19.64–25.44 mm
  Mean22.79±1.74 mm
Junction
  AP15.42±1.59 mm11.80–18.46 mm
  RU14.75±1.39 mm12.32–17.31 mm
Length
  Ventral11.81±1.57 mm9.09–15.44 mm
  Dorsal11.77±1.41 mm9.77–14.80 mm
  Radial11.89±1.52 mm8.34–15.04 mm
  Ulnar11.72±1.42 mm9.67–15.42 mm
  Mean11.80±1.19 mm
Curvature radius8.76±2.50 mm5.57–17.67 mm
Angle
  Ventral153.1°±9.1°134°–171.9°
  Dorsal164.8°±3.8°155.1°–170.3°
  Radial154°±8.6°136.4°–168.7°
  Ulna163.5°±8.1°147.4°–162.2°

Measurement Results in Comparable Studies

StudyRange, mm
Head–Diaphysis Angle
Head SizeLength
Current study23.1522.4411.8911.72153°–165°
Beredjiklian et al723221212N/A
Popovic et al822.821.8N/A9.93N/A
Itamura et al1022.320.99.8N/AN/A
Swieszkowski et al1123.3622.26N/AN/AN/A
Captier et al1221.62115.3N/A167°–169°