Cloud-based Preoperative Planning for Total Hip Arthroplasty: A Study of Accuracy, Efficiency, and Compliance

Joseph D. Maratt, MD; Ramesh C. Srinivasan, MD; William J. Dahl, MD; Peter L. Schilling, MD; Andrew G. Urquhart, MD

Abstract

As digital radiography becomes more prevalent, several systems for digital preoperative planning have become available. The purpose of this study was to evaluate the accuracy and efficiency of an inexpensive, cloud-based digital templating system, which is comparable with acetate templating. However, cloud-based templating is substantially faster and more convenient than acetate templating or locally installed software. Although this is a practical solution for this particular medical application, regulatory changes are necessary before the tremendous advantages of cloud-based storage and computing can be realized in medical research and clinical practice.

Drs Maratt, Srinivasan, Dahl, Schilling, and Urquhart are from the Department of Orthopaedic Surgery, University of Michigan, Ann Arbor, Michigan.

Drs Maratt, Srinivasan, Dahl, Schilling, and Urquhart have no relevant financial relationships to disclose.

Correspondence should be addressed to: Andrew G. Urquhart, MD, Department of Orthopaedic Surgery, University of Michigan, 2912 Taubman Center, 1500 E Medical Center Dr, Ann Arbor, MI 48109 (andrewu@med.umich.edu).

Abstract

As digital radiography becomes more prevalent, several systems for digital preoperative planning have become available. The purpose of this study was to evaluate the accuracy and efficiency of an inexpensive, cloud-based digital templating system, which is comparable with acetate templating. However, cloud-based templating is substantially faster and more convenient than acetate templating or locally installed software. Although this is a practical solution for this particular medical application, regulatory changes are necessary before the tremendous advantages of cloud-based storage and computing can be realized in medical research and clinical practice.

Drs Maratt, Srinivasan, Dahl, Schilling, and Urquhart are from the Department of Orthopaedic Surgery, University of Michigan, Ann Arbor, Michigan.

Drs Maratt, Srinivasan, Dahl, Schilling, and Urquhart have no relevant financial relationships to disclose.

Correspondence should be addressed to: Andrew G. Urquhart, MD, Department of Orthopaedic Surgery, University of Michigan, 2912 Taubman Center, 1500 E Medical Center Dr, Ann Arbor, MI 48109 (andrewu@med.umich.edu).

Total hip arthroplasty (THA) is a reliable treatment for hip arthritis. Successful THA consistently relieves pain, improves function, and restores biomechanics of the hip. However, improper component selection or malposition causing excessive wear, limb-length discrepancy, fracture, or dislocation can lead to early failure or patient dissatisfaction. Therefore, preoperative planning is an important step for achieving a successful outcome after THA.1–4

Prior studies examining the accuracy of templating have shown more than 90% accuracy to within 2 sizes of the implant. However, the exact size of the component was predicted in only 42% to 50% of cases for the acetabular component and 68% for the femoral component.4,5

As digital radiography has become more prevalent, tools for digital preoperative planning have become available for use by the treating surgeon. On-screen evaluation of digital imaging does not affect the accuracy of interpretation.6–8 However, when printed digital radiographs are combined with acetate overlay templating, correct component size is predicted accurately only 69% of the time.9

Studies investigating the accuracy of digital templating have produced varied results. Gonzalez Della Valle et al10 showed consistently better accuracy with conventional templating compared with digital templating in a study of 69 patients undergoing cemented THA. In contrast, Iorio et al11 reported that the mean errors did not differ significantly and concluded that both acetate and digital templating can accurately predict the size of THA implants. For these studies, digital templating was performed using software packages installed on site.10,11

In the current study, the authors used cloud-based software for digital templating (HipCAT; University of Michigan, Ann Arbor, Michigan). Cloud computing is a relatively new model in information technology that provides fee-for-service access to computing resources via the Internet.12 It avoids the high costs associated with traditional in-house computing infrastructure deployment and expansion. Instead, computing is provided as a metered service and resembles the service provided by a public utility company that supplies gas or electricity.12–14 Advocates of this model highlight its low cost, ease of adoption, scalability, and performance.14,15 However, the model continues to be challenged by concerns over privacy and security.12,16–18

Many of the advantages of cloud computing are applicable to digital templating systems. This digital templating system did not require the purchase of additional hardware or software or software deployment. It also avoided the expense associated with systems integration by providing a utility to import images into the cloud saved from a picture archiving and communication systems (PACS) viewer (McKesson Corp, San Francisco, California). All functions required for templating, reviewing, and uploading images are accessible through a Web browser. Privacy and security concerns were addressed with Health Insurance Portability and Accountability (HIPAA)-compliant encryption implemented on the Web browser–based client and the cloud-based service.

The purpose of this study was to measure the accuracy and efficiency of cloud-based digital and acetate templating for preoperative planning of THA. The authors also address the regulatory compliance issues related to cloud computing.

Materials and Methods

A retrospective review of 20 primary uncemented THAs performed between September and December 2010 under the direction of 1 adult reconstruction fellowship-trained surgeon (A.G.U.) was performed. Patients who underwent prior THA on the contralateral hip, had metal implants on the operative side, or had nonstandard imaging were excluded from the study.

Standardized techniques were applied when preoperative radiography was obtained. Anteroposterior pelvis radiographs were downloaded from the PACS system. The downloaded images were deidentified and then printed on analog film or uploaded as an encrypted file into the cloud-based templating software. The images were then calibrated in the templating system based on the scaling overlay used by the PACS system and a magnification marker when present. In a separate session, the operative report was reviewed to determine the actual implanted acetabular and femoral components for later statistical comparison.

Measurement of Templating Accuracy

One attending (A.G.U.) and 2 senior orthopedic surgical residents (R.C.S., W.J.D) performed preimplant assessments of hip implant sizing on the 20 cases with traditional acetate templates and with digital templating software. This was done in 2 sessions spaced 2 weeks apart. Acetate templating was performed according to a previously described standardized method.3,17 Digital templating was performed according to the digital templating software protocol, allowing marking and measurement similar to acetate templating (Figure).

Anteroposterior view of the pelvis after preoperative planning with digital templates.

Figure: Anteroposterior view of the pelvis after preoperative planning with digital templates.

The template results for both methods were compared with the actual acetabular and femoral components implanted.

Measurement of Templating Timing

The preparation of each preoperative plan by both methods was individually timed by an observer (J.D.M.) and recorded. The timer began when each film was hung on the lightbox or the image was loaded. The timer was stopped when the participant indicated satisfaction with the templating.

Statistical Analysis

PASW Statistics version 19 software (IBM, Armonk, New York) was used for statistical analysis. Interrater reliability was measured with intraclass correlation coefficients. Two-tailed 1-sample t tests were used to assess the accuracy of templating by both of these methods for each participant. The time required to template by each method was also recorded. Differences in time for each method of templating were analyzed using paired t tests.

Results

Measurement of Templating Accuracy

Digital templating predicted correct acetabular component size within 2 sizes in 96% of cases and within 1 size in 73% of cases. Acetate overlay templating was similarly accurate to 2 sizes in 86% of cases and to 1 size in 63% of cases. Digital templating predicted correct femoral component size within 2 sizes in 93% of cases and 1 size in 75% of cases. Acetate templating was accurate to 2 sizes in 98% of cases and 1 size in 93% of cases. Table 1 shows the distribution of error by size for each component.

Distribution of Acetabular and Femoral Component Error

Table 1: Distribution of Acetabular and Femoral Component Error

Moderate and comparable interrater reliability was found with digital and acetate templating of the acetabular component (Table 2). However, slightly lower interrater reliability was found with acetate templating of the femoral component compared with the digital method.

Interrater Reliability as Assessed With Intraclass Correlation Coefficientsa

Table 2: Interrater Reliability as Assessed With Intraclass Correlation Coefficients

Analysis of pooled measurement data showed significant underprediction of femoral and acetabular component size with digital and acetate overlay templating (Tables 3, 4). Digital measurements resulted in greater underprediction of femoral component size (negative error values, P<.005). Acetate measurements resulted in greater underprediction of acetabular component size (negative error values, P<.05). However, the standard deviation of the estimation error with both digital and acetate templating is reflective of the number of cases in which the components were oversized.

Acetabular Component Size Accuracy

Table 3: Acetabular Component Size Accuracy

Femoral Component Size Accuracy

Table 4: Femoral Component Size Accuracy

Measurement of Templating Timing

Analysis of the time required to complete the tasks revealed that acetate templating on average took 2.07 times longer than digital templating. Although some variation existed in the amount of time required for completion of the task by each participant, acetate templating was consistently slower (1.47 to 2.53 times longer to complete the task depending on the participant). On average, digital templating required 64 seconds per case and acetate templating required 133 seconds per case (P<.005). Results for each participant are shown in Table 5.

Time Required for Completion of Templating

Table 5: Time Required for Completion of Templating

Discussion

In the current study, digital and acetate templating were comparably accurate for planning THA. Digital and acetate templating predicted components to within 2 sizes in the vast majority of cases (>90% of the time). Digital and acetate templating also showed comparable interrater reliability for the selection of each component.

The results are comparable with prior studies of the accuracy of digital and acetate preoperative templating with uncemented THA. Iorio et al11 reported that components could be predicted within 1 size in 60% to 74% of cases with digital templating and in 77% to 78% of cases with acetate templating. The current authors’ results showed prediction of components within 1 size in 73% to 75% of cases with digital templating and 58% to 93% of cases with acetate templating. These results are also comparable with prior studies examining the accuracy of acetate templating in planning uncemented THA; these studies reported exact prediction of component size 42% to 68% of the time and within 2 sizes greater than 90% of the time.4,5

In contrast, Gonzalez Della Valle et al10 reported that analog templating was more predictable and precise than digital templating. In their study, analog templating predicted components within 1 size in 97% to 98% of cases, whereas digital templating from scanned analog radiographs predicted components within 1 size in 81% to 94% of cases. However, in their study, a cemented technique was used for placement of the femoral stem.10 Multiple investigators have reported that prediction of cemented components is more reliable in preoperative planning than prediction of uncemented components.10,18–20 Thus, if a cemented technique for the femoral stem is used, analog templating may be more accurate than digital templating.

To the authors’ knowledge, no other studies to date have examined the efficiency of analog and digital templating. This study showed that surgeons needed approximately half the time to prepare a preoperative plan with digital templating when compared with analog templating.

With the increasing prevalence of digital radiography and use of PACS in the hospital and office setting, digital templating may be a more efficient and cost-effective method for preoperative planning. Potential benefits to surgeon and hospital include reduced expense from film production, secure storage, improved access, and increased efficiency of templating. Another advantage of digital templating is the ease of error correction. As a result, digital templating is a valuable tool for teaching and discussion of cases, especially in a teaching hospital setting.

A cloud-based system confers additional benefits beyond traditional on-site PACS and digital templating systems. Cloud-based systems for electronic health records and medical imaging software may lower financial barriers and promote acquisition.12 However, questions remain regarding the security and privacy of information stored on these systems.

Use of these systems in medical applications must comply with the privacy and security rules of the HIPAA Act and of the Health Information Technology for Economic and Clinical Health Act. Federal law requires that third parties handling protected health information enter into a Business Associate Contract with the client that stipulates adherence to privacy guidelines. Liability now extends to the business associates for violations of the security provisions.12,21

Development of these systems with a focus on enhanced security is sensible when storing sensitive medical data. The digital templating system used in this study encrypts all medical data on the local computer prior to transmission and storage. With this approach, in the event of a security breach of the cloud, the data remain unintelligible without each user’s encryption key.15,22 These approaches to security and legislative policy make the use of cloud computing feasible in the medical industry.

Conclusion

Preoperative planning for THA with cloud-based digital templating is comparable with acetate templating in accuracy but offers the additional benefits of efficiency, cost savings, and workflow improvement. Cloud-based medical software can be safely implemented in compliance with federal privacy and security regulations. However, individual physicians and institutions should be judicious consumers of nascent technology with potential security and privacy pitfalls.

References

  1. Berry DJ, Harmsen WS, Cabanela ME, Morrey BF. Twenty-five-year survivorship of two thousand consecutive primary Charnley total hip replacements: factors affecting survivorship of acetabular and femoral components. J Bone Joint Surg Am. 2002; 84(2):171–177.
  2. Charnley J. Total hip replacement by low-friction arthroplasty. Clin Orthop Relat Res. 1970; 72:7–21.
  3. Della Valle AG, Padgett DE, Salvati EA. Preoperative planning for primary total hip arthroplasty. J Am Acad Orthop Surg. 2005; 13(7):455–462.
  4. Unnanuntana A, Wagner D, Goodman SB. The accuracy of preoperative templating in cementless total hip arthroplasty. J Arthroplasty. 2009; 24(2):180–186. doi:10.1016/j.arth.2007.10.032 [CrossRef]
  5. Carter LW, Stovall DO, Young TR. Determination of accuracy of preoperative templating of noncemented femoral prostheses. J Arthroplasty. 1995; 10(4):507–513. doi:10.1016/S0883-5403(05)80153-6 [CrossRef]
  6. Roos JE, Chilla B, Zanetti M, et al. MRI of meniscal lesions: soft-copy (PACS) and hard-copy evaluation versus reviewer experience. AJR Am J Roentgenol. 2006; 186(3):786–790. doi:10.2214/AJR.04.1853 [CrossRef]
  7. Shields T. PACS: past, present and future. Radiol Technol. 2010; 81(5):491–498.
  8. Eklund K, Jonsson K, Lindblom G, et al. Are digital images good enough? A comparative study of conventional film-screen vs digital radiographs on printed images of total hip replacement. Eur Radiol. 2004; 14(5):865–869. doi:10.1007/s00330-003-2126-y [CrossRef]
  9. Conn KS, Clarke MT, Hallett JP. A simple guide to determine the magnification of radiographs and to improve the accuracy of preoperative templating. J Bone Joint Surg Br. 2002; 84(2):269–272. doi:10.1302/0301-620X.84B2.12599 [CrossRef]
  10. Gonzalez Della Valle A, Comba F, Taveras N, Salvati EA. The utility and precision of analogue and digital preoperative planning for total hip arthroplasty. Int Orthop. 2008; 32(3):289–294. doi:10.1007/s00264-006-0317-2 [CrossRef]
  11. Iorio R, Siegel J, Specht LM, Tilzey JF, Hartman A, Healy WL. A comparison of acetate vs digital templating for preoperative planning of total hip arthroplasty: is digital templating accurate and safe?J Arthroplasty. 2009; 24(2):175–179. doi:10.1016/j.arth.2007.11.019 [CrossRef]
  12. Schweitzer EJ. Reconciliation of the cloud computing model with US federal electronic health record regulations. J Am Med Inform Assoc. 2011; 19(2)161–165. doi:10.1136/amiajnl-2011-000162 [CrossRef]
  13. Katzan H Jr, . On an ontological view of cloud computing. J Service Sci. 2010; 3(1):1–6.
  14. National Institute of Standards and Technology. Cloud Computing. http://www.nist.gov/itl/cloud/index.cfm. Accessed August 11, 2011.
  15. Armbrust M, Fox A, Griffith R, et al. A view of cloud computing. Commun ACM. 2010; 53(4):50–58. doi:10.1145/1721654.1721672 [CrossRef]
  16. Katzan H Jr, . On the privacy of cloud computing. Int J Manag Info Syst. 2010; 14(2):1–12.
  17. Callaghan JJ, Rosenberg AG, Rubash HE. The Adult Hip. 2nd ed. Philadephia, PA: Lippincott Williams & Wilkins; 2007.
  18. Gonzalez Della Valle A, Slullitel G, Piccaluga F, Salvati EA. The precision and usefulness of preoperative planning for cemented and hybrid primary total hip arthroplasty. J Arthroplasty. 2005; 20(1):51–58. doi:10.1016/j.arth.2004.04.016 [CrossRef]
  19. Knight JL, Atwater RD. Preoperative planning for total hip arthroplasty. Quantitating its utility and precision. J Arthroplasty. 1992; 7(suppl):403–409. doi:10.1016/S0883-5403(07)80031-3 [CrossRef]
  20. The B, Diercks RL, van Ooijen PM, Van Horn JR. Comparison of analog and digital preoperative planning in total hip and knee arthroplasties. A prospective study of 173 hips and 65 total knees. Acta Orthop. 2005; 76(1):78–84. doi:10.1080/00016470510030364 [CrossRef]
  21. U.S. Department of Health & Human Services. Business Associates. http://www.hhs.gov/ocr/privacy/hipaa/understanding/coveredentities/businessassociates.html. Accessed August 11, 2011.
  22. Amazon Web Services. Creating HIPAA-Compliant Medical Data Applications With AWS. http://aws.amazon.com/about-aws/whats-new/2009/04/06/whitepaper-hipaa/. Accessed August 11, 2011.

Distribution of Acetabular and Femoral Component Error

Error (sizes)a Acetabular Component, %
Femoral Component, %
Digital Acetate Digital Acetate
−3+ 3.33 11.67 6.67 1.67
−2 18.33 20.00 15.00 5.00
−1 15.00 16.67 20.00 26.67
0 38.33 30.00 38.33 56.67
1 20.00 11.67 16.67 10.00
2 5.00 8.33 3.33 0.00
3+ 0.00 1.67 0.00 0.00

Interrater Reliability as Assessed With Intraclass Correlation Coefficientsa

Method Acetabular Component Femoral Component
Digital 0.52 0.42
Acetate 0.52 0.34

Acetabular Component Size Accuracy

Participant Digital Errora P Acetate Errora P
1 −0.65±1.35 <.05 −0.90±1.92 <.05
2 −0.05±1.05 .83 −0.25±1.33 .41
3 −0.25±1.25 .38 −0.70±1.45 <.05
Combined −0.31±1.23 <.05 −0.61±1.58 <.05

Femoral Component Size Accuracy

Participant Digital Errora P Acetate Errora P
1 −1.10±1.25 <.005 −0.65±0.81 <.05
2 0.25±0.78 .17 1.10±0.55 .42
3 −0.55±1.23 .06 −0.40±0.83 <.05
Combined −0.46±1.22 <.005 −0.31±0.79 <.05

Time Required for Completion of Templating

Participant Digital Time, sec Acetate Time, sec P
1 63.45±17.66 93.00±18.02 <.005
2 66.65±13.12 168.80±79.03 <.005
3 82.95±28.10 157.80±27.38 <.005
Combined 64.35±15.47 133.25±1.27 <.005

10.3928/01477447-20120725-05

Sign up to receive

Journal E-contents