Journal of Refractive Surgery

Letters to the Editor Free

Optimal Pocket Depth for Corneal Inlays

Majid Moshirfar, MD; Brian D. Walker, BS; Steve H. Linn, OD; Orry C. Birdsong, MD; Phillip C. Hoopes Jr., MD

To the Editor:

Changes in refraction following small-aperture corneal inlay implantation (KAMRA; AcuFocus, Irvine, CA) for the treatment of presbyopia are mainly due to hyperopic shift.1–3 A possible explanation for this is a wound healing response postoperatively, with keratocyte activation and stromal thickening, resulting in changed corneal topography.2,3

The AcuFocus guidelines for implantation suggest that the intrastromal pocket be created at a depth of 200 to 250 µm while maintaining a residual stromal bed of 250 µm.4 We postulated that shallow pocket depth is associated with greater wound healing response due to keratocyte populations being denser in the anterior stroma.5 Thus, increasing pocket depth would lead to less wound healing response and subsequent hyperopic shift. Recently, we compared the change in manifest refraction spherical equivalent (MRSE) and uncorrected near visual acuity (UNVA) over time for patients who received small-aperture corneal inlays at intrastromal pocket depths of 200 to 249 and 250 to 300 µm. This study was approved by the HDR Research Review Board (Draper, Utah) and all data were collected in accordance with the tenets of the Declaration of Helsinki from patients who provided informed consent.

Seventy-nine patients had small-aperture corneal inlays implanted between May 2015 and April 2017. Fifty-two inlays (66%) were implanted at a depth of 200 to 249 µm and 27 inlays (34%) were implanted at a depth of 250 to 300 µm. The average preoperative MRSE for the two groups was −0.45 and −0.49 diopters (D), respectively. MRSE and UNVA were reported for 3 and 12 months postoperatively.

Both groups maintained good uncorrected distance visual acuity, with 89% seeing 20/40 or better at 12 months compared to 93% preoperatively. At 3 months, UNVA of Jaeger 5 (J5) or better was 90% for 250 µm or greater and 81% for less than 250 µm. At 12 months, UNVA of J5 or better was 89% for 250 µm or greater and 77% for less than 250 µm. At 12 months postoperatively, average MRSE changed +0.45 and −0.17 D for the less than 250 µm and 250 µm or greater groups, respectively (Figure 1). Hyperopic shift occurred in the shallow group and did not in the deeper group.

Change in manifest spherical equivalent refraction (MRSE) from preoperative baseline.

Figure 1.

Change in manifest spherical equivalent refraction (MRSE) from preoperative baseline.

These preliminary data reflect that the ideal intrastromal pocket for the KAMRA inlay is deeper (250 to 300 µm) than the suggested 200 to 250 µm for the best visual outcomes. This may be because increased pocket depth leads to less wound healing response and can be further evaluated with the aid of confocal microscopy in future studies. However, it is important to maintain the integrity of the residual stromal bed while increasing the depth of the KAMRA pocket. Further study on intrastromal pocket depth may be warranted based on this preliminary data.

Majid Moshirfar, MD

Brian D. Walker, BS

Steve H. Linn, OD

Orry C. Birdsong, MD

Phillip C. Hoopes, Jr., MD

Salt Lake City, Utah

References

  1. United States Food and Drug Administration. Summary of Safety and Effectiveness Data (SSED) (P120023). Available at: https://www.accessdata.fda.gov/cdrh_docs/pdf12/P120023B.pdf
  2. Dexl AK, Jell G, Strohmaner C, et al. Long-term outcomes after monocular corneal inlay implantation for the surgical compensation of presbyopia. J Cataract Refract Surg. 2015;41:566–575. doi:10.1016/j.jcrs.2014.05.051 [CrossRef]
  3. Abbouda A, Javaloy J, Alió J. Confocal microscopy evaluation of the corneal response following AcuFocus KAMRA inlay implantation. J Refract Surg. 2014;30:172–178. doi:10.3928/1081597X-20140217-04 [CrossRef]
  4. AcuFocus. KAMRA® Inlay Professional Use Information. (n.d.). Available at: https://www.accessdata.fda.gov/cdrh_docs/pdf12/p120023d.pdf
  5. Patel SV, Mclaren JW, Hodge DO, et al. Normal human density and corneal thickness measurement by using confocal microscopy in vivo. Invest Ophthalmol Vis Sci. 2001;42:333–339.
Authors

The authors have no financial or proprietary interest in the materials presented herein.

10.3928/1081597X-20180209-01