Evaluation and Treatment of Painful Total Hip Arthroplasties With Modular Metal Taper Junctions

R. Michael Meneghini, MD; Nadim J. Hallab, PhD; Joshua J. Jacobs, MD

  • Orthopedics
  • May 2012 - Volume 35 · Issue 5: 386-391
  • DOI: 10.3928/01477447-20120426-03
Rights and Permissions

Abstract

Modern primary total hip arthroplasty femoral components have evolved to include modular necks. Subsequently, the additional taper junction provides another interface as a potential source for mechanically assisted crevice corrosion, which is a complex process involving fretting and crevice corrosion. Furthermore, it is becoming evident that an adverse local tissue reaction may result in some patients due to the mechanically assisted crevice corrosion. This article details the clinical, radiographic, and laboratory evaluation of patients with these components who present with persistent pain. The relevant surgical strategies and techniques to address this pathology in symptomatic patients are addressed.

Dr Meneghini is from Indiana University Health Physicians, Department of Orthopaedic Surgery, Indiana University School of Medicine, Indianapolis, Indiana; and Drs Hallab and Jacobs are from the Department of Orthopaedic Surgery, Rush University Medical Center, Chicago, Illinois.

Dr Meneghini recieves funds from and is a consultant for Stryker Orthopaedics and has patents with Nemcomed. Dr Hallab has no relevant financial relationships to disclose. Dr Jacobs is a consultant for Johnson & Johnson, Medtronic Sofamor Danek, Smith & Nephew, and Zimmer, and receives royalties from Taylor and Francis.

Correspondence should be addressed to: R. Michael Meneghini, MD, Indiana University Health Physicians, Department of Orthopaedic Surgery, Indiana University School of Medicine, 13100 136th St, Ste 2000, Fishers, IN 46037 (rm_meneghini@yahoo.com).

Modern primary total hip arthroplasty (THA) femoral components have evolved to include those with modular necks, which provide the surgeon with more options to reproduce or correct the patient’s anatomy and hip biomechanics. However, with the additional taper junction, the modular neck femoral components provide an additional interface that may be a source of mechanically assisted crevice corrosion, a complex process involving fretting corrosion and crevice corrosion.1–7 Reported to occur at the metal taper junction between cobalt-chromium (CoCr) femoral heads and metal stems, mechanically assisted crevice corrosion can similarly occur at the metal taper junction between the modular neck and stem in such designs. This can occur in titanium–titanium, CoCr–titanium, and CoCr–CoCr taper junctions.1–7

It has been reported that elevated serum metal levels and an adverse local tissue reaction may occur with metal-on-metal bearings, causing premature failure.8–24 For the purposes of this report, adverse local tissue reactions include soft tissue responses, such as aseptic lymphocyte-dominated vasculitis-associated lesions, periprosthetic osteolysis, and pseudotumors. A similar reaction and potentially premature failure of THA has been reported to occur in rare cases of patients with modular-neck femoral components.3

This article discusses the clinical presentation, evaluation, and workup of patients who present with persistent pain after successful THA with a metal taper junction suspected of mechanically assisted crevice corrosion. This will provide clinicians with basic guidelines—but only at our current level of understanding based on the few reported cases—for the potential clinical presentation, evaluation, histopathology, and surgical treatment of such patients.

Clinical Presentation

Patients with a persistently painful THA containing a modular neck primary femoral component (Figure 1) who lack an evident etiology should be considered for a potential adverse local tissue reaction to mechanically assisted crevice corrosion at the modular taper junction. Patients will typically present with pain in the anterior, lateral, or posterior hip. In a severe case of pseudotumor, patients may feel a large mass or fluid collection, raising the suspicion for significant local soft tissue destruction. Patients may have pain at rest, but more often have pain with weight bearing. Physical examination may also reveal pain with hip motion and joint loading.

Radiograph of a painful total hip replacement with a modular neck.

Figure 1: Radiograph of a painful total hip replacement with a modular neck.

Evaluation for Infection

Anecdotal and reported cases document superimposed bacterial infection in the face of clinical adverse local tissue reaction from metal-on-metal articulations.25 Evidence also exists that alteration of the local pH may create an environment that facilitates metal corrosion, which suggests the pH alteration in occult periprosthetic infection may be a precursor to corrosion in metal taper junctions.26 Further confounding the issue is the reported incidence of patients with an adverse local tissue reaction who present with a clinical picture mimicking infection, including elevated serum inflammatory markers.20

The initial workup of a painful THA should be a screening of the erythrocyte sedimentation rate and C-reactive protein; if either or both are elevated, a hip aspiration is essential to rule out infection as the source of pain. In patients with a modular taper junction who do not have evidence for a diagnosis for infection, a workup for the possibility of adverse local tissue reaction from metal junction mechanically assisted crevice corrosion is recommended.

In patients with metal-on-metal bearings and an adverse local tissue reaction, the intra-articular fluid is typically substantial in volume, turbid, and brownish or grey in color.9 Despite this frequently described appearance, aspiration analysis has not yet been proven to be specific to failed implants due to metal-on-metal bearings or modular taper junctions.

Serum Metal Levels

Mechanically assisted crevice corrosion from modular taper junctions may result in adverse local tissue reactions via elevated metal concentrations. Therefore, patients presenting with a painful THA of unknown etiology and a modular taper junction should obtain serum metal studies. It has been reported that chromium ion levels >17 μg/L and cobalt ion levels >19 μg/L were associated with elevated joint fluid ion levels and metallosis.28

Recently, Kwon et al29 reported metal-on-metal resurfacing arthroplasties using ultrasound/magnetic resonance imaging and serum/hip aspirate cobalt and chromium measurements. Pseudotumors found in 7 patients (4%) were associated with higher cobalt and chromium levels. The median serum cobalt and chromium levels in women with bilateral arthroplasty in the pseudotumor group were 9.0 and 12.0 μg/L, respectively, but only 2.9 and 3.2 μg/L, respectively, in similar patients without pseudotumor formation. These data suggest that pseudotumors may be associated with increased wear generated from metal-on-metal articulations.29

The Medicines and Healthcare products Regulatory Agency suggested a cutoff blood level of 7 parts per billion, yet this was shown to have a specificity of 89% and sensitivity 52% for preoperatively detecting an unexplained failed metal-on-metal THA.30 It was reported the optimal cutoff for the maximum cobalt or chromium level was 4.97 parts per billion and had sensitivity 63% and specificity 86%; it was concluded that blood metal ions have the ability to separate failed and well-functioning hip replacements.30

However, interlaboratory variability can exist in metal measurements due to differences in instrumentation, specimen collection protocols, matrix (whole blood or serum), and analytic methodology.31 In addition, the levels may be confounded by the presence of other implants containing metal and differences in renal function. Although an understanding of metal levels and their correlation with adverse local tissue reactions in THA continues to evolve, the above values are attempts at providing some guidance to the clinician evaluating a patient with a painful THA suspected of an adverse local tissue reaction with modular taper junctions.

Metal Hypersensitivity

Speculation exists that a metal allergy may contribute to implant loosening. However, a clear connection is not found between the incidence of metal sensitivity and implant duration, infection, reason for removal, or pain. At this time, it is unclear whether metal sensitivity causes implant loosening or vice versa. Diagnostic testing in the form of either patch testing or an in vitro lymphocyte transformation test may be considered in addition to metal levels to address the patients’ level of metal exposure and their potential for a hypersensitivity response.33 Thomas et al35 investigated 16 patients with a failed metal-on-metal bearing hip arthroplasty due to lymphocyte dominated periprosthetic inflammation. In 13 (81%) patients, systemic metal sensitivity was found based on patch testing, the in vitro lymphocyte transformation test, or both.35

Radiographic Analysis

Radiographic analysis of painful THAs should always include plain radiographs to evaluate for loosening and osteolysis and for the assessment of implant position. Particular attention should be paid to the medial calcar, trochanteric, and periacetabular regions for cortical lucency or erosion. In many cases of adverse local tissue reaction from mechanically assisted crevice corrosion or metal-on-metal bearings, the plain radiographs are normal and show no evidence of bony erosion (Figure 1). In patients with persistent pain, elevated serum metal ion levels, and normal plain radiographs, advanced imaging studies are recommended.

Ultrasound has been used to evaluate fluid collections and masses associated with pseudotumors or seen adjacent to the implants. However, more recently, computed tomography and magnetic resonance imaging with metal artifact reduction sequence have been useful to delineate more subtle soft tissue findings associated with adverse local tissue reaction around metal-on-metal bearings.36 Magnetic resonance imaging detects osteolysis and complications in the periprosthetic soft tissues, such as wear-induced synovitis, periprosthetic fluid collections, neurovascular compression, and quality of the muscle and tendons.37–40 If one is considering an MRI, detailed dialogue with radiologists knowledgeable in the metal artifact reduction techniques is necessary to optimize the information available from these advanced imaging techniques.41,42

Periarticular Histopathology

Patients with an adverse local tissue reaction to local metal debris from mechanically assisted crevice corrosion may demonstrate similar perivascular pathologic characteristics as those reported in patients with metal-on-metal bearings. Intraoperative hip fluid is typically described as brownish or grey in color.9 The characteristic histological features typically consist of diffuse and perivascular infiltrates of T and B lymphocytes and plasma cells, high endothelial venules, massive fibrin exudation, accumulation of macrophages with drop-like inclusions, and infiltrates of eosinophilic granulocytes and necrosis (Figure 2). Also, metal particles are uncommon and are not a requisite for the diagnosis of an adverse local tissue reaction.8,9,16,20,24,35

Tissue section (hematoxylin and eosin stain, ×200 original magnification) of peri-articular tissue from a patient with crevice corrosion showing fibroconnective tissue with necrosis and viable areas of engorged blood vessels and diffuse lymphocytic infiltrate.

Figure 2: Tissue section (hematoxylin and eosin stain, ×200 original magnification) of peri-articular tissue from a patient with crevice corrosion showing fibroconnective tissue with necrosis and viable areas of engorged blood vessels and diffuse lymphocytic infiltrate.

Recommended Surgical Treatment

Although no specific clinical data exist to guide surgeons in the surgical treatment of patients who have a painful THA with modular taper junction mechanically assisted crevice corrosion at the head–neck or modular neck–stem junction, certain tenets and strategies should be used. First, although difficult and stressful for the surgeon and patient, it is not recommended that observation without surgical intervention be used for any significant time. Adverse local tissue reactions have been reported in patients with mechanically assisted crevice corrosion of metal taper junctions.

In severe cases, these adverse local tissue reactions can render abductors completely absent or via mass effect cause neurovascular compromise of surrounding nerves or vessels. Therefore, surgical intervention should be strongly considered once the diagnosis of an adverse local tissue reaction from mechanically assisted crevice corrosion has been confirmed. As always, this is done after a detailed discussion with the patients explaining the risks of observation vs operative intervention.

Once the decision to surgically intervene has been made, the method of revision reconstruction should use the principle of minimizing any modular metal taper junctions because the presumption is either that the patient’s immune system played a role in the development of an adverse local tissue reaction via hypersensitivity or that the patient’s local biology created an environment supportive of mechanically assisted crevice corrosion for reasons that are yet unclear. Therefore, acetabular components can typically be retained if well-fixed and in acceptable position. The acetabular liner should be composed of highly cross-linked polyethylene, although ceramic liners may be used as well.

If a femoral component with a modular neck was used, then consideration should be given to removal and replacement with a femoral component that does not use a modular neck. If mechanically assisted crevice corrosion damage is limited to the male portion of the taper, one may consider retaining a well-fixed stem and exchanging the damaged modular neck; however, data are not yet available to determine the likelihood or incidence of mechanically assisted crevice corrosion recurring at this interface. Therefore, the surgeon should weigh the risks associated with removal of a well-fixed femoral component with the uncertainty associated with a modular neck exchange. However, if visible damage is observed to the femoral stem portion of the taper junction, strong consideration should be given to removal of the femoral component.

In further support of minimizing metal modular interfaces in patients who have adverse local tissue reactions, if a metal femoral head was used in the original reconstruction, consideration should be given to replacement with a modern ceramic femoral head. It has been shown that ceramic-on-metal modular junctions have less mechanically assisted crevice corrosion than metal-on-metal modular junctions.2 However, if damage exists to an existing metal taper, it is recommended to use a ceramic head with a metal sleeve insert to minimize the chance of ceramic head fracture. Fortunately, the majority of studies that report surgical treatment of adverse local tissue reaction due to metal-on-metal bearings by conversion to an alternative bearing such as ceramic-on-ceramic or metal-on-polyethylene have reported resolution of pain and symptoms in the majority of patients (Figure 3).43

Successful revision of the patient in Figure 1 to a cementless femoral component without a modular neck, using a ceramic-on-polyethylene bearing to minimize any potential for recurrence of and adverse local tissue reaction due to mechanically assisted crevice corrosion.

Figure 3: Successful revision of the patient in Figure 1 to a cementless femoral component without a modular neck, using a ceramic-on-polyethylene bearing to minimize any potential for recurrence of and adverse local tissue reaction due to mechanically assisted crevice corrosion.

Intraoperative Findings

Evidence of mechanically assisted crevice corrosion of the modular taper junction (either at the neck–head or neck–stem junction) is visualized by irregular black material on the surface of the metal contained within the junction (Figure 4). Furthermore, the black material is typically associated with and near surface irregularities on the metal taper surface in contact with the opposite metal surface, consistent with crevice corrosion.3 Adverse local tissue reactions may also create a substantial amount of intra-articular joint fluid, as is typically seen preoperatively on MRI, and is reported to be a brownish or grey color with a turbid consistency.9

Modular neck with significant mechanically assisted crevice corrosion visible on the male taper junction.

Figure 4: Modular neck with significant mechanically assisted crevice corrosion visible on the male taper junction.

Although this has been described, to the authors’ knowledge the intra-articular fluid characteristics have not been consistently characterized macroscopically or with respect to laboratory analysis and cell count at the time of this publication. Local soft tissue necrosis has been seen in adverse local tissue reaction from metal debris and can be extensive to include the pseudocapsule, the abductor musculature, and tendinous insertion onto the greater trochanter (Figure 5), with necrotic bone seen in the most severe cases. In addition, medial femoral neck or calcar resorption, aseptic loosening, and osteolysis from metal debris have been observed.

Intraoperative photograph of patient in Figure 1 showing peri-articular necrosis and significant soft tissue damage secondary consistent with an adverse local tissue reaction due to mechanically assisted crevice corrosion.

Figure 5: Intraoperative photograph of patient in Figure 1 showing peri-articular necrosis and significant soft tissue damage secondary consistent with an adverse local tissue reaction due to mechanically assisted crevice corrosion.

References

  1. Brown SA, Flemming CA, Kawalec JS, et al. Fretting corrosion accelerates crevice corrosion of modular hip tapers. J Appl Biomater. 1995; 6(1):19–26. doi:10.1002/jab.770060104 [CrossRef]
  2. Hallab NJ, Messina C, Skipor A, Jacobs JJ. Differences in the fretting corrosion of metal-metal and ceramic-metal modular junctions of total hip replacements. J Orthop Res. 2004; 22(2):250–259. doi:10.1016/S0736-0266(03)00186-4 [CrossRef]
  3. Kop AM, Swarts E. Corrosion of a hip stem with a modular neck taper junction: a retrieval study of 16 cases. J Arthroplasty. 2009; 24(7):1019–1023. doi:10.1016/j.arth.2008.09.009 [CrossRef]
  4. Reclaru L, Eschler PY, Lerf R, Blatter A. Electrochemical corrosion and metal ion release from Co-Cr-Mo prosthesis with titanium plasma spray coating. Biomaterials. 2005; 26(23):4747–4756. doi:10.1016/j.biomaterials.2005.01.004 [CrossRef]
  5. Vendittoli PA, Roy A, Mottard S, Girard J, Lusignan D, Lavigne M. Metal ion release from bearing wear and corrosion with 28 mm and large-diameter metal-on-metal bearing articulations: a follow-up study. J Bone Joint Surg Br. 2010; 92(1):12–19. doi:10.1302/0301-620X.92B1.22226 [CrossRef]
  6. Kop AM, Keogh C, Swarts E. Proximal component modularity in THA-at what cost? An implant retrieval study [published online ahead of print November 3, 2011]. Clin Orthop Relat Res. In press.
  7. Gilbert JL, Buckley CA, Jacobs JJ. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy coupling. J Biomed Mater Res. 1993; 27(12):1533–1544. doi:10.1002/jbm.820271210 [CrossRef]
  8. Bolland BJ, Culliford DJ, Langton DJ, Millington JP, Arden NK, Latham JM. High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: clinical, radiological and retrieval analysis. J Bone Joint Surg Br. 2011; 93(5):608–615. doi:10.1302/0301-620X.93B5.26309 [CrossRef]
  9. Browne JA, Bechtold CD, Berry DJ, Hanssen AD, Lewallen DG. Failed metal-on-metal hip arthroplasties: a spectrum of clinical presentations and operative findings. Clin Orthop Relat Res. 2010; 468(9):2313–2320. doi:10.1007/s11999-010-1419-0 [CrossRef]
  10. Donell ST, Darrah C, Nolan JF, et al. Early failure of the Ultima metal-on-metal total hip replacement in the presence of normal plain radiographs. J Bone Joint Surg Br. 2010; 92(11):1501–1508. doi:10.1302/0301-620X.92B11.24504 [CrossRef]
  11. Ebramzadeh E, Campbell PA, Takamura KM, et al. Failure modes of 433 metal-on-metal hip implants: how, why, and wear. Orthop Clin North Am. 2011; 42(2):241–250. doi:10.1016/j.ocl.2011.01.001 [CrossRef]
  12. Hasegawa M, Yoshida K, Wakabayashi H, Sudo A. Pseudotumor with dominant b-lymphocyte infiltration after metal-on-metal total hip arthroplasty with a modular cup [published online ahead of print Jun 30, 2011]. J Arthroplasty. 2012; 27(3):493.e5–7. doi:10.1016/j.arth.2011.05.005 [CrossRef]
  13. Hutt JR, Busch C, Hughes RA. Failure of a metal on metal hip prostheses presenting as a destructive soft tissue mass due to ALVAL. Rheumatol Int. 2011; 31(10):1401–1402. doi:10.1007/s00296-010-1668-y [CrossRef]
  14. Iwamoto T, Ikari K, Momohara S. Pseudotumor from a metal-on-metal hip. J Rheumatol. 2011; 38(10):2265. doi:10.3899/jrheum.110806 [CrossRef]
  15. Klapperich C, Graham J, Pruitt L, Ries MD. Failure of a metal-on-metal total hip arthroplasty from progressive osteolysis. J Arthroplasty. 1999; 14(7):877–881. doi:10.1016/S0883-5403(99)90042-6 [CrossRef]
  16. Kwon YM, Thomas P, Summer B, et al. Lymphocyte proliferation responses in patients with pseudotumors following metal-on-metal hip resurfacing arthroplasty. J Orthop Res. 2010; 28(4):444–450.
  17. Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br. 2010; 92(1):38–46. doi:10.1302/0301-620X.92B1.22770 [CrossRef]
  18. Maezawa K, Nozawa M, Matsuda K, Yuasa T, Shitoto K, Kurosawa H. Early failure of modern metal-on-metal total hip arthroplasty using a Wagner standard cup. J Arthroplasty. 2006; 21(4):522–526. doi:10.1016/j.arth.2005.07.012 [CrossRef]
  19. Maurer-Ertl W, Friesenbichler J, Liegl-Atzwanger B, Kuerzl G, Windhager R, Leithner A. Noninflammatory pseudotumor simulating venous thrombosis after metal-on-metal hip resurfacing. Orthopedics. 2011; 34(10):e678–e681.
  20. Mikhael MM, Hanssen AD, Sierra RJ. Failure of metal-on-metal total hip arthroplasty mimicking hip infection. A report of two cases. J Bone Joint Surg Am. 2009; 91(2):443–446. doi:10.2106/JBJS.H.00603 [CrossRef]
  21. Ollivere B, Darrah C, Barker T, Nolan J, Porteous MJ. Early clinical failure of the Birmingham metal-on-metal hip resurfacing is associated with metallosis and soft-tissue necrosis. J Bone Joint Surg Br. 2009; 91(8):1025–1030. doi:10.1302/0301-620X.91B8.21701 [CrossRef]
  22. Park YS, Moon YW, Lim SJ, Yang JM, Ahn G, Choi YL. Early osteolysis following second-generation metal-on-metal hip replacement. J Bone Joint Surg Am. 2005; 87(7):1515–1521. doi:10.2106/JBJS.D.02641 [CrossRef]
  23. Shahrdar C. Pseudotumor in large-diameter metal-on-metal total hip articulation. J Arthroplasty. 2011; 26(4):665 e621–663. doi:10.1016/j.arth.2010.05.022 [CrossRef]
  24. Willert HG, Buchhorn GH, Fayyazi A, et al. Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study. J Bone Joint Surg Am. 2005; 87(1):28–36. doi:10.2106/JBJS.A.02039pp [CrossRef]
  25. Watters TS, Eward WC, Hallows RK, Dodd LG, Wellman SS, Bolognesi MP. Pseudotumor with superimposed periprosthetic infection following metal-on-metal total hip arthroplasty: a case report. J Bone Joint Surg Am. 2010; 92(7):1666–1669. doi:10.2106/JBJS.I.01208 [CrossRef]
  26. Merritt K, Brown SA. Effect of proteins and pH on fretting corrosion and metal ion release. J Biomed Mater Res. 1988; 22(2):111–120. doi:10.1002/jbm.820220204 [CrossRef]
  27. Schinsky MF, Della Valle CJ, Sporer SM, Paprosky WG. Perioperative testing for joint infection in patients undergoing revision total hip arthroplasty. J Bone Joint Surg Am. 2008; 90(9):1869–1875. doi:10.2106/JBJS.G.01255 [CrossRef]
  28. De Smet K, De Haan R, Calistri A, et al. Metal ion measurement as a diagnostic tool to identify problems with metal-on-metal hip resurfacing. J Bone Joint Surg Am. 2008; 90(suppl 4):202–208. doi:10.2106/JBJS.H.00672 [CrossRef]
  29. Kwon YM, Ostlere SJ, McLardy-Smith P, Athanasou NA, Gill HS, Murray DW. “Asymptomatic” pseudotumors after metal-on-metal hip resurfacing arthroplasty: prevalence and metal ion study. J Arthroplasty”. 2011; 26(4):511–518. doi:10.1016/j.arth.2010.05.030 [CrossRef]
  30. Hart AJ, Sabah SA, Bandi AS, et al. Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement. J Bone Joint Surg Br. 2011; 93(10):1308–1313. doi:10.1302/0301-620X.93B10.26249 [CrossRef]
  31. Clark M, Prentice J, Hoggard N, Jacobs JJ, Stockley I, Wilkinson JM. Effect of analysis laboratory on metal levels after MoMHR and potential impact on patient management and interpretation of research datasets. Paper presented at: Orthopaedic Research Society Annual Meeting. ; February 4–7, 2012. ; San Francisco, CA. .
  32. Jacobs JJ, Gilbert JL, Urban RM. Corrosion of metal orthopaedic implants. J Bone Joint Surg Am. 1998; 80(2):268–282.
  33. Jacobs JJ, Urban RM, Hallab NJ, Skipor AK, Fischer A, Wimmer MA. Metal-on-metal bearing surfaces. J Am Acad Orthop Surg. 2009; 17(2):69–76.
  34. Hallab NJ. Lymphocyte transformation testing for quantifying metal-implant-related hypersensitivity responses. Dermatitis. 2004; 15(2):82–90. doi:10.2310/6620.2004.03054 [CrossRef]
  35. Thomas P, Braathen LR, Dorig M, et al. Increased metal allergy in patients with failed metal-on-metal hip arthroplasty and peri-implant T-lymphocytic inflammation. Allergy. 2009; 64(8):1157–1165. doi:10.1111/j.1398-9995.2009.01966.x [CrossRef]
  36. Cahir JG, Toms AP, Marshall TJ, Wimhurst J, Nolan J. CT and MRI of hip arthroplasty. Clin Radiol. 2007; 62(12):1163–1171. doi:10.1016/j.crad.2007.04.018 [CrossRef]
  37. Hayter CL, Potter HG, Su EP. Imaging of metal-on-metal hip resurfacing. Orthop Clin North Am. 2011; 42(2):195–205. doi:10.1016/j.ocl.2010.12.006 [CrossRef]
  38. Mistry A, Cahir J, Donell ST, Nolan J, Toms AP. MRI of asymptomatic patients with metal-on-metal and polyethylene-on-metal total hip arthroplasties. Clin Radiol. 2011; 66(6):540–545. doi:10.1016/j.crad.2011.02.002 [CrossRef]
  39. Sabah SA, Mitchell AW, Henckel J, Sandison A, Skinner JA, Hart AJ. Magnetic resonance imaging findings in painful metal-on-metal hips: a prospective study. J Arthroplasty. 2011; 26(1):71–76. doi:10.1016/j.arth.2009.11.008 [CrossRef]
  40. Toms AP, Marshall TJ, Cahir J, et al. MRI of early symptomatic metal-on-metal total hip arthroplasty: a retrospective review of radiological findings in 20 hips. Clin Radiol. 2008; 63(1):49–58. doi:10.1016/j.crad.2007.07.012 [CrossRef]
  41. Hayter CL, Koff MF, Shah P, Koch KM, Miller TT, Potter HG. MRI after arthroplasty: comparison of MAVRIC and conventional fast spin-echo techniques. AJR Am J Roentgenol. 2011; 197(3):W405–W411. doi:10.2214/AJR.11.6659 [CrossRef]
  42. Toms AP, Smith-Bateman C, Malcolm PN, Cahir J, Graves M. Optimization of metal artefact reduction (MAR) sequences for MRI of total hip prostheses. Clin Radiol. 2010; 65(6):447–452. doi:10.1016/j.crad.2009.12.014 [CrossRef]
  43. Rajpura A, Porter ML, Gambhir AK, Freemont AJ, Board TN. Clinical experience of revision of metal on metal hip arthroplasty for aseptic lymphocyte dominated vasculitis associated lesions (ALVAL). Hip Int. 2011; 21(1):43–51. doi:10.5301/HIP.2011.6276 [CrossRef]
AUTHORS

Dr Meneghini is from Indiana University Health Physicians, Department of Orthopaedic Surgery, Indiana University School of Medicine, Indianapolis, Indiana; and Drs Hallab and Jacobs are from the Department of Orthopaedic Surgery, Rush University Medical Center, Chicago, Illinois.

Dr Meneghini recieves funds from and is a consultant for Stryker Orthopaedics and has patents with Nemcomed. Dr Hallab has no relevant financial relationships to disclose. Dr Jacobs is a consultant for Johnson & Johnson, Medtronic Sofamor Danek, Smith & Nephew, and Zimmer, and receives royalties from Taylor and Francis.

Correspondence should be addressed to: R. Michael Meneghini, MD, Indiana University Health Physicians, Department of Orthopaedic Surgery, Indiana University School of Medicine, 13100 136th St, Ste 2000, Fishers, IN 46037 (rm_meneghini@yahoo.com).

doi: 10.3928/01477447-20120426-03

Healio is intended for health care provider use and all comments will be posted at the discretion of the editors. We reserve the right not to post any comments with unsolicited information about medical devices or other products. At no time will Healio be used for medical advice to patients.